1, the correct use of electric constant temperature incubator (1) in strict accordance with the instructions for use, the correct use of electric constant temperature incubator;
(2) Choose a reasonable temperature field within the working area of ​​the temperature field;
(3) When using a constant temperature incubator, ensure that the constant temperature is long enough to maintain a constant temperature field;
(4) In the steady state, it ensures high accuracy of its temperature field.
2. How to improve the accuracy of the use of an electrothermal incubator Before the measurement, we introduce several concepts about the temperature field.
Incubator work area: In a steady state, it conforms to the temperature uniformity of the incubator and the temperature fluctuations in the area. Incubator temperature uniformity: In the steady state, the difference between the highest temperature and the lowest temperature in the constant temperature incubator working area. Incubator temperature fluctuation: In the steady state, the incubator work area within a certain time interval, the range of temperature changes. Incubator Temperature Deviation: In steady state, the incubator shows the difference between the temperature and the actual temperature at the geometric center of the workspace.
The following step by step how to improve the use of electrothermal temperature incubator temperature field accuracy.
1 First of all, check the primary components and the temperature display instrument of the electrothermic incubator, and give the correction values ​​of the corresponding measurement temperature points. This eliminates the combined errors introduced by the primary and secondary temperature instruments.
2 Due to the difference between the location of the temperature measuring element and the working geometric center of the working area in the temperature field, the system error naturally occurs. We measured and corrected the difference between the temperature at the geometric center of the work area and the temperature at which the primary element was located, thus avoiding errors due to the temperature measurement element not being in the geometric center.
3 When conducting the temperature field test, the error caused by the standard device should also be taken into account, ie the error caused by the standard device should also be corrected.
4 The electrothermal incubator is mainly a temperature indicator, so the accuracy of temperature control also directly affects the accuracy of the temperature field. Even if the temperature indicating error of the temperature indicating instrument in the incubator is 0, if the temperature control accuracy is not high, the temperature control point deviates from the actual temperature also causes the temperature field to run out of control. Therefore, setpoint errors for each temperature control point should be given and corrected.
5 In summary, to improve the accuracy of the use of electrothermal incubators, the above error sources should be corrected.
Thorough Hardened screw HPT series is imported alloy steel. After high-temperature quenching and tempering treatment, the hardness of the round rod reaches to HRC60. Under the different types of rough & finishing grinding progress, it meets the requirements of high precision for all kinds of products and prolongs the service life of the screw in different applications.
HPT series is suitable for plastic material with 25%-65% glass fiber, with wear resistance, corrosion resistance, high temperature resistance, high pressure resistance, and high speed resistance property. It is suitable for fully electrical injection molding machine with high speed, high processing requirements. According to the different customers demands, the PVD coating and Chrome plating is applied to improve the wear resistance and corrosion resistance.
Welcome to purchase!
Thorough Hardened Screw,Case Hardened Bolts,Hardened Screw,Machine Screw
ZHEJIANG HUAYE PLASTICS MACHINERY CO.,LTD , https://www.huaye-group.com